NemuStim and ECE50
Functional Check of Vocal Folds via Neuromuscular Stimulation
The NemuStim stimulation device with the ECE50 endoscopic cap electrode serves to assess the motility of the vocal folds using electrical stimulation. This innovative procedure is already being utilized in leading ENT clinics and delivers quick results about unilateral and bilateral vocal fold motility disorders. It can be an outpatient treatment or performed in the operating room.
The diagnostic is gentle yet efficient: The endoscopic cap electrode is placed on a compatible video endoscope and inserted into the mouth without requiring additional invasive procedures.
Quick and Efficient
Electric neuromuscular stimulation with NemuStim delivers quick results to support diagnostics.
Versatile and Convenient
NemuStim can be used in both inpatient and outpatient settings and offers comprehensive compatibility.
Interested in Our Products?
If you are interested in using NemuStim and/or ECE50 or have questions about these products, please contact us directly via E-Mail:
We look forward to hearing from you!
Further Information & Documents for Download
Why MED-EL:
A Trusted Partner
For more than 35 years, MED-EL has been a trusted partner and innovation leader in hearing implants. We’re here to work together with you, and we’re committed to providing outstanding service and support for our professional partners.
With the most advanced cochlear implant systems, we offer the best hearing experience for your patients and the best clinical experience for you.
Technical Data
NemuStim Neuromuscular Stimulation Device
- Weight: ~200 g
- Up to 5 hours of continuous battery life in stimulation mode (3-hour charging time from empty)
- Acoustic feedback during stimulation
- Optional display background light
- Galvanic isolation, no direct current
- Single-pulse and burst modes
- Low-frequency, biphasic rectangular pulses
- Duration of pulses: 0.1–2.1 ms
- Output current: 0.1–10 mA
- Output voltage: 25 V at 1,000 Ω
- Frequency range: 1–100 Hz
- 2 stimulation cable connectors, Ø 2 mm (yellow, white)
- 3 feedback electrode connectors, Ø 1.5 mm (black, black, green)
- EMG measurement range: ±300 μV
ECE50 Endoscopic Cap Electrode
- Approved for adults only
- Sterile, single-use product
- Material: medical silicone, titanium nitride electrode contacts
- Length of the electrode tongue: 50 mm
- Ø 2 mm touch-protected contacts
- Electrical pulse ranges with amplitudes from 0.1–10 mA (at 1–60 Hz) and pulse duration of 0.2–2.1 ms
- Compatible with Ø 5.9 mm video endoscopes (with or without working channel)
- Compatible with the handheld NemuStim or other stimulators that cover the parameter range of ECE50
Please find detailed information and specific conditions for use for each product in the respective instructions for use. Not all products, indications, and features shown are available in all areas.
* Age limitations of connected electrodes apply